How to usability test AI-enabled clinical software
When a clinical tool is built on a model, the interesting failures are rarely crashes. They are a clinician who accepts a wrong suggestion because the screen looked confident, or one who stops reading a flag because it fired too often. A usability test of an AI tool has to be designed to produce those moments on purpose, and an accuracy score will never show them.
The short answer
Test the person and the model together. Give participants realistic cases in which the output is clearly right, clearly wrong and borderline, and watch what they do with each. Record use errors and close calls, then ask why. If the tool is regulated as software as a medical device, this belongs inside your usability engineering process alongside the usual critical-task testing.
Model accuracy and usability are separate questions. A model that is right 95 per cent of the time can still cause harm if people cannot tell which 5 per cent to doubt. A model that is rarely right can still be useful if the interface makes its limits obvious.
The use errors that are specific to AI tools
Ordinary software has misclicks and misread labels. AI-enabled tools add a few failure patterns of their own:
- Over-reliance. The user accepts an output without checking it, including when it is wrong. This is often called automation bias, and it grows when the tool is usually right.
- Under-reliance. The user ignores or overrides correct outputs, because earlier ones were wrong, because alerts fire too often, or because they cannot see why the tool said what it said.
- Misreading the output. A probability read as a certainty, a risk band read as a diagnosis, a confidence figure read as accuracy. How you display uncertainty is part of the design.
- Working outside the tool's scope. Using it on a patient group, image type or setting it was not built for, because nothing on screen said not to.
- Losing the thread. A change in the model's output after an update that the user does not notice, so they act on old assumptions.
- Workarounds. Copying outputs into another system, or keeping a paper list alongside, which creates new transcription errors.
Designing tasks that surface them
The task set matters more than the number of sessions. A few principles:
- Include cases where the tool is wrong. If every test case has a correct output, you are only measuring whether people can read the screen. Seed wrong and borderline outputs, written with a clinician who can say what a realistic miss looks like.
- Use realistic data. Messy records, missing values and conflicting notes. Tidy test data hides the situations where clinicians lean on the tool most.
- Add realistic pressure. A ward round, an interrupting phone call, a queue of cases. A calm, unlimited-time session tends to produce better decisions than real use does.
- Do not say the tool might be wrong. If you warn participants, you test the warning. Brief them as they would be briefed in real life.
- Test the explanation. If the tool shows reasons, evidence or confidence, check whether people use them to catch errors or just treat them as decoration.
Who should take part
Representative means matched on the things that change behaviour: role, specialty, seniority and experience with similar tools. A consultant and a junior doctor may read the same output differently, and a sceptical user and an enthusiast will certainly behave differently. Include both.
For a regulated product, each distinct user group is counted separately. Formative rounds use five to eight per group, and summative validation uses at least 15 per group. Our guide to how many participants medical device usability testing needs goes through how to define the groups.
We recruit clinicians through our professional panel and clinical networks, so the people in the room do the job you are building for.
What to record
Satisfaction scores tell you whether people liked the tool. They do not tell you whether it is safe. For each critical task, record:
- Whether the participant reached the right decision, and whether they would have without the tool
- Use errors and close calls, with what the participant said and did around them
- What they looked at on the screen before deciding, and what they never looked at
- A root-cause interview afterwards: why did you accept, override or ignore that?
The root-cause conversation is where the design changes come from. It is also what a regulator will look for when you explain how you judged a use error acceptable or fixed it.
When to retest
A model can change without any screen changing. A retrain, a new data source or a threshold adjustment can alter what users see, and so how they behave. Treat a change in output behaviour as a trigger to review your usability evidence, not only a change to the interface.
You do not need to repeat a full summative study every time. A short written assessment of which critical tasks are affected, followed by a targeted retest of those tasks, is usually proportionate. Agree the approach with your regulatory lead.
What it costs
A formative round of eight sessions on regulated health software starts from £9,750, including recruitment and participant payments. Programmes of several rounds start from £16,000. Summative validation is scoped per user group. See SaMD usability testing for details.
Questions AI product teams ask us
Do we need to test the model itself? Model performance is validated separately, with data. Usability testing looks at what people do with the output.
Can we use synthetic users or AI personas? Not for this. Whether a clinician over-trusts a recommendation depends on training, workload and habit, which a model of a user cannot supply. See why AI personas cannot replace real user research.
Is this only for tools regulated as medical devices? No. The same approach helps any clinical or health tool that uses a model. If it is regulated, the findings also feed your evidence file. Our AI product testing page covers the wider case.
We fully recognise the effort your team invested in recruitment, moderation, and analysis, and we genuinely appreciate the quality of the discussions and reporting.
- Usability problems in an insulin device surfaced before manufacturing locked anything in, with human factors evidence supporting the regulatory submission.
- An NHS patient portal team got a prioritised, evidenced list of adoption barriers, and the highest-leverage fix wasn't the one they expected.
- Patient involvement changed a trial's primary endpoint before the protocol was finalised, cited by the ethics committee as a model for evidencing PPIE.
Trusted PPI and PPIE delivery partner to the NIHR HealthTech Research Centre in Accelerated Surgical Care.
Building an AI tool for clinicians or patients?
Tell us what it recommends, who relies on it and where you are in development. We will help you design cases that show how people really use it, and come back with a plan and a fixed price.
Talk to us about your toolWe usually respond within one working day.
Prefer to talk it through first?
Book a free 20-minute call. No obligation, no sales pitch. We'll tell you honestly whether research is worth it for your decision, and what it would cost.