Software becomes a medical device because of what you say it is for, not because of how it is built. A plain spreadsheet can be one. A clever AI model might not be. Once it is one, user testing stops being a good habit and becomes part of the evidence you have to produce. This guide covers how to tell, and what testing follows, from a UK point of view.

The short answer

The short answer

If your software is intended to diagnose, monitor, predict, treat or relieve a condition, or to inform decisions about any of those, it is likely to be a medical device. In Great Britain that means it falls under the MHRA and the UK medical device regulations. Software that only stores records, handles bookings or provides general wellbeing information usually does not.

If it is a device, you need to show that the people it is intended for can use it safely. For software, that usually means a usability engineering process following IEC 62366-1, with formative testing while you build and summative testing before you file.

This is general information from a research practice, not regulatory advice. Your regulatory lead or UK Approved Body decides your classification and what evidence you need.

Intended purpose

Intended purpose decides it

The question is not what the software is. It is what you, the manufacturer, claim it does. The claim lives in your website copy, your sales deck, your instructions and your marketing, and regulators read all of it.

Some pairs that show the difference:

  • A symptom diary that stores what a patient types is a record. The same diary that analyses the entries and tells a clinician which patients to see first is making a clinical suggestion.
  • A step counter with a wellbeing score is a lifestyle app. The same sensor data used to flag a possible arrhythmia is monitoring a condition.
  • A dose table in a PDF is reference material. A calculator that takes a patient's weight and returns a dose is informing a treatment decision.
  • A scheduling tool for clinics is administration. A triage tool that sorts patients by urgency is not.

Changing your wording can move a product out of scope. It can also leave you with a product that does not do what you wanted. Decide the purpose on purpose, write it down and make sure your marketing matches.

UK route

What the UK route looks like

In Great Britain, devices are regulated by the MHRA under the UK medical device regulations. Most manufacturers place a UKCA mark or, for a transition period, a CE mark that is still accepted in Great Britain. Northern Ireland follows different rules. The transition dates and the rules for classifying software have both been changing, so check the current position with the MHRA or a regulatory specialist before you plan around them.

Two practical points affect your user testing:

  • Class changes the evidence you are expected to hold. Higher-class software normally goes through a UK Approved Body, which will want to see the usability engineering file. Lower-class software is often self-declared, but the manufacturer still carries the duty to show the software is safe to use.
  • The MHRA publishes human factors guidance for medical devices, and IEC 62366-1 is the usability engineering standard most teams follow. Following a recognised standard is usually the simplest way to show you have done what a regulator would expect.
What testing it triggers

What user testing the process asks for

IEC 62366-1 sets out a process. The parts that involve real people are these:

  • Use specification. Who uses the software, for what, where and under what conditions. Good ones come from observing real use, not from a workshop of colleagues.
  • Use-related risk analysis. Which tasks could lead to harm if done wrong, and how. The tasks with the highest risk become your critical tasks.
  • Formative evaluation. Testing with representative users during design to find use problems you can still fix. Five to eight people per user group per round is usual, over several rounds.
  • Summative evaluation. A final study showing representative users can complete the critical tasks without unacceptable risk. A common expectation is at least 15 participants per distinct user group. Our guide to how many participants medical device usability testing needs explains why.

For software, the interface includes the screens, the notifications, the in-app help, the onboarding and any instructions that ship with it. All of that is part of what you are testing.

Selling to the NHS

The extra layer when the NHS is the customer

Regulatory approval is one hurdle. NHS buyers add their own.

  • DCB0129 and DCB0160. These are the NHS clinical risk management standards, one for manufacturers and one for the organisations deploying the software. Both rely on a hazard log, and what you learn from user testing feeds it.
  • The Digital Technology Assessment Criteria (DTAC). The usability and accessibility section asks how you have involved users and whether your product meets accessibility standards. Testing with disabled users is a practical way to answer it, and we can add it to the same study.

A single, well-planned study with the right participants can serve all of these, as long as it is designed that way from the start.

Software is different

Where software testing differs from hardware

  • It changes after release. Each update is a chance to introduce a new use error. You need a way to judge which changes need retesting and which do not, and to write the reasoning down.
  • It runs on other people's equipment. Patients use their own phones, with their own settings, text sizes and screen readers. Clinicians use hospital computers with locked-down browsers. Both matter.
  • It sits inside a workflow. A clinical tool is used between interruptions, alongside other systems. Testing it in a quiet room with unlimited time will miss errors that happen on a busy ward.
  • Remote testing is often fine. Software-only products can usually be tested remotely, which is cheaper and recruits more widely. Agree the format with your regulatory team first.
Cost

What it costs

A formative round of eight sessions on regulated health software starts from £9,750, including recruitment and participant payments. Multi-round programmes start from £16,000. Summative validation is scoped once the user groups and critical tasks are known. Details are on our SaMD usability testing page.

Common questions

Questions software teams ask us

Does an AI model make it a medical device? No. The intended purpose does. An AI tool that summarises meeting notes is not. One that suggests a diagnosis is likely to be.

Is a wellbeing or mental health app a medical device? Only if it claims a medical purpose, such as treating or monitoring a condition. General wellbeing features usually are not, but the line is easy to cross with marketing wording.

Can we start testing before we know our classification? Yes, and you should. Formative testing improves the product whatever the answer, and the findings feed your risk analysis when you do classify.

Do you decide whether we need a summative study? No. That is for your regulatory lead or Approved Body. We help you plan and run it once you know.

We fully recognise the effort your team invested in recruitment, moderation, and analysis, and we genuinely appreciate the quality of the discussions and reporting.

UK Operations Manager Medicsen

Trusted PPI and PPIE delivery partner to the NIHR HealthTech Research Centre in Accelerated Surgical Care.

Reviewing findings on a phone alongside printed research materials

Testing software that may be a medical device?

Tell us what it does and who uses it. We will help you work out the user groups and critical tasks, and come back with a plan and a fixed price, usually within one working day.

Talk to us about your software

We usually respond within one working day.

Prefer to talk it through first?

Book a free 20-minute call. No obligation, no sales pitch. We'll tell you honestly whether research is worth it for your decision, and what it would cost.